Product Specifications

 Components

ThermaCor Rapid Infuser 10 mL/hour to 1,200 mL/minute
Quick-Loading Cassette One-Latch to engage 4 air sensors, 3 heat sensors, 3 pressure sensors and 1 cassette sensor
Patient Line – 6’  
Dual Patient Line – 6’  
Extension Line – 3’  
3 Spike Inflow Set  
Large Fluid Reservoir  

 Operating Characteristics

Slow Infusion    10 mL/hour to 1,200 mL/hour
Rapid Infusion    20 mL/minute to 1,200 mL/minute
Bolus Infusion    100, 250, 500 or 1,000 mL of fluid
Bolus Selectable Rate    20mL/minute to 1200 mL/minute at tolerance indicated above
Recirc Mode    300mL/min for 5 minutes
Total infusion amounts    up to 99,999 mL
Maximum infusion pressure    100 or 300 mmHg
Nominal fluid inlet temperature    20°C
Nominal fluid outlet temperature w/ Heater On    37±2°C at 1,000 mL/minute
Output Temperature    Set to 40±2°C (Displays temperature in 1° increments)
Protection against air infusion    Prevents infusion of air ³50 mL at max rated flow
Time required for priming   <  60 seconds

 Environmental Characteristics

Operating Temperature 5°C to 40°C
Relative humidity 20% to 90%
Altitude        0 to 10,000 feet
EMC (radiated/immunity) CISPR11 Group 1 Level B/IEC 1000-3 Level 2
Mechanical noise <65 dBA in normal operation @ 1 meter

 Storage and Transportation Characteristics

Pump    -40°C to 70°C, 10% to 100% condensing Humidity
Cassette and Tubing    -40°C to 70°C, 10% to 100% condensing Humidity

 Physical Characteristics

Size     W8.75” x H16.5” x D13”
Weight (without Disposable)

    22 lbs

IV pole mount    Accepts IV pole 0.75 to 1.5” in diameter.  IV Pole should be capable
   of safely supporting at least 30 lb
Power Cord

Conductor: 3-14AWG 14 x 30 strand, 9.7mm OD

Cord type/temp:  SJT/105deg C

Power Rating: 1875 Watts 15A @ 125V

Hospital Grade plug

 Power Source

AC Input Voltage    1380 VA Max 100-240 VAC
Operating Frequency    50/60Hz

Battery backup – Type

- Operating cell capacity

   Lithium ion battery pack

250 mL/hour for 2 hours

or 1,000 mL/minute for 30 minutes*